User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: ME 26D/U DATE: 16-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 26 NUMBER OF LINES: 115 CONFIGURATION: Fluke 5100B ()B CONFIGURATION: Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F 1.002 DISP Ensure that the Test Instrument has Warmed Up. 1.002 DISP Switch the Test Instrument Power Off. 1.003 DISP Zero the meter movement using the center screw. 1.004 DISP Set the Test Instrument SELECTOR switch to + 1.005 HEAD {Volts DC and Linearity Tests} 1.006 DISP Set the RANGE switch to 1V 1.007 DISP Connect the Test Instrument DC PROBE (black) and 1.007 DISP COMMON (alligator clip) Test Leads together and adjust 1.007 DISP the ZERO ADJ for a meter indication of 0. 1.008 DISP Disconnect the Test Leads 1.008 DISP Connect the DC Test Lead to the 'Binding Posts to Double 1.008 DISP Banana Plug Adapter' by placing the probe tip in the 1.008 DISP small side hole under the red plastic head and 1.008 DISP tightening down the plastic head. The probe should be 1.008 DISP attached from the right hand side facing the red plastic 1.008 DISP head at the top. Connect the COMMON alligator clip into 1.008 DISP the black binding post. 1.009 DISP The following tests use the DC, COMMON leads and 1.009 DISP Adapter for connection to the Test Interface Panel. 1.010 ASK- R 1.011 5100 1 0.1V 3/ 2W 2.001 5100 1 0.5V 3/ 2W 3.001 5100 1 1V 3/ 2W 4.001 ASK+ R 4.002 5100 3 3V 3/ 2W 5.001 5100 10 10V 3/ 2W 6.001 5100 30 30V 3/ 2W 7.001 5100 100 100V 3/ 2W 8.001 5100 300 300V 3/ 2W 9.001 5205 1000 1000V 3/ 2W 10.001 DISP Set SELECTOR switch to - 10.002 5205 1000 -1000V 3/ 2W 11.001 DISP Remove the connections from the Test Interface Panel. 11.002 HEAD {Volts AC Tests} 11.003 DISP Set Test Instrument SELECTOR switch to AC 11.003 DISP Set RANGE switch to 1V 11.004 DISP Disconnect the Test Leads from the binding posts of the 11.004 DISP Adapter. Connect the Test Instrument AC PROBE and AC 11.004 DISP GROUND leads together and adjust the AC ZERO knob 11.004 DISP for a meter indication of 0. 11.005 DISP Disconnect the Test Leads. 11.005 DISP Connect the AC PROBE and AC GROUND leads to the 'Binding 11.005 DISP Posts to Double Banana Plug Adapter' in the same manner 11.005 DISP described previously. 11.006 DISP The following tests use the AC PROBE and Adapter for 11.006 DISP connection to the Test Interface Panel. 11.007 ASK- R 11.008 5100 1 1V 3/ 400H 2W 12.001 ASK+ R 12.002 5100 3 3V 3/ 400H 2W 13.001 5100 10 10V 3/ 400H 2W 14.001 5100 30 30V 3/ 400H 2W 15.001 5100 100 100V 3/ 400H 2W 16.001 5100 300 300V 3/ 400H 2W 17.001 HEAD {Frequency Response Tests} 17.002 DISP Disconnect the AC PROBE and Adapter from the TI Panel. 17.003 DISP Connect the 'Binding Posts to BNC Plug Adapter' on to 17.003 DISP the Adapter now connected to the AC PROBE. 17.004 DISP Connect the 50Ohm Feedthrough to the BNC Plug connected 17.004 DISP to the AC PROBE. 17.005 DISP The following tests use the AC PROBE and Adapters for 17.005 DISP connections to the Test Interface Panel. 17.006 WIDE 1 1V 12% 20H 2W 18.001 WIDE 1 1V 12% 1MH 2W 19.001 WIDE 1 1V 12% 10MH 2W 20.001 HEAD {Resistance Tests} 20.002 ASK- R 20.003 DISP Remove the connections from the Test Interface Panel. 20.004 DISP Disconnect the Test Leads from the Adapter 20.004 DISP Set the Test Instrument SELECTOR switch to OHMS 20.005 DISP Set UUT Range to {RX10} 20.006 DISP Ensure that the probes are not touching one another 20.006 DISP and set the OHMS ADJ knob for an indication of infinity 20.006 DISP on the Test Instrument. Short the meter leads together 20.006 DISP at the Test Interface panel and set the ZERO ADJ for a 20.006 DISP meter indication of zero. Remove the short. 20.007 ASK- W 20.008 IEEE [@0]L-AB,L+7, 20.009 DISP Move the connection at the Test Interface Panel from 20.009 DISP the Multifunction Input Terminals to the 20.009 DISP Resistance Source Terminals. Also use a banana to 20.009 DISP banana jumper to connect the Resistance source low 20.009 DISP to the green GND terminal on the Test Interface Panel. 20.010 RESF 100Z 5% 2W 21.001 DISP Set UUT Range to {RX100} 21.002 DISP Ensure that the probes are not touching one another 21.002 DISP and set the OHMS ADJ knob for an indication of infinity 21.002 DISP on the Test Instrument. Short the meter leads 21.002 DISP together at the Test Interface Panel and set the ZERO 21.002 DISP ADJ knob for a meter indication of zero. 21.002 DISP Remove the short. 21.003 RESF 1000Z 5% 2W 22.001 DISP Set UUT Range to {RX1K} 22.002 RESF 10KZ 5% 2W 23.001 DISP Set UUT Range to {RX10K} 23.002 RESF 100KZ 5% 2W 24.001 DISP Set UUT Range to {RX100K} 24.002 RESF 1000KZ 5% 2W 25.001 DISP Set UUT Range to {RX1M} 25.002 RESF 10MZ 5% 2W 26.001 END